EAST AND WEST GC-4000A Gaz Kromotografi Cihazı

EAST AND WEST GC-4000A  Gaz Kromotografi Cihazı

GC-4000A Series is well suited for analyzing of a wide range of chemicals as gases, liquids or gas-liquid mixtures with boiling points less than 400℃. This versatile instrument can be used in industries such as the oil, coal mining, chemical, environmental protection, textile, agricultural, medicinal, and sanitation industry. 

Technical Specifications

TCD

FID

ECD

FPD

NPD

Sensitivity

S≥5000
mv.ml/mg (benzene)

Minimum Detectable Quantity

M≤1×10-11
g/sec
(hexadecane)

M ≤1×10-13
g/ml(r-666)

Mp≤2×10-12
g/s(1605)
Ms≤5×10
-11
g/s( thiophene )

MN≤5×10-13
g/s
Mp≤1×10-12
g/s

Noise

≤0.1mv

≤5×0-14A

≤0.01mv

≤2×10-12A

≤2×10-13A

Linear Dynamic Range

106

103

103

103

Drift

1×10-14
A/30min

0.02mv/30min

3×10-13
A/30min

2.5×10-12 A/h

 

. We offer five types of detectors. Each one is specifically designed to fit your needs.

√ Flame Ionization Detector (FID): Used for analyzing organic compound requiring high sensitivity.
√ Thermal Conductivity Detector (TCD): Used for analyzing ambient temperature gases, water, and small hydrocarbon compounds.
√ Flame Photometric Detector (FPD): Used for analyzing sulfurous or phosphorous compounds.
√ Electron Capture Detector (ECD): Used for detecting pesticides or chlorides through Electron Capture
√ Nitrogen-Phosphorus Detector (NPD): Used for trace analysis of nitrides or phosphides

2. Easy installion and maintence
3. Large column oven that be simultaneously installed with three columns.
4. Efficient design. Instrument can be installed with up to three types of detectors at once, three independent electrical circuit systems and three signal outputs.
5. Flexible packed column and capillary column system.
6. Stable temperature control system: Temperature programming of up to 8 ramps; individually controlled temperatures for each sample injector and detector; Multiple sensors to prevent over-heating.
7. Improved pressure stabilization and multiple gas path systems: a single, dual or triple gas path; Injector and valve sample injection system; Back flushing and pre-cutting devices; Methane conversion oven.

Our Complete Set

• Main unit
• Data processing workstation: a PC (optional), a printer (optional), workstation software and interface board
• Gas sources/generators: air, hydrogen gas and high purity nitrogen gas (optional)
• Chromatographic column(s) (please specify)
• Spare parts

The following generators can be chosen for GC:
GH-200(G-101) High purity hydrogen generator
GH-100(G-101A) High purity hydrogen generator
GA-100(G-103) High purity air generator

GC-4000A Series Data Sheet (Please click here)

GC-4000A Gas Chromatographs (Large Screen Liquid Crystal Display)

Technical Specifications


1. Column Oven
Dimensions: 300×300×200mm
Operating Temperatures: Room Temperature +5℃ - 400℃
Temperature Precision: 1%
Overheating protection
Precision of the Temperature Control: ±0.1℃
2. Large LCD screen for displaying temperature, external events, operational conditions, and other information
3. Automatic rear door control for over-heating protection.
4. Temperature Programming
Number of Programmable Ramps: 8

Temperature Range: Room Temperature +5℃ - 400℃
Temperature Ramp Rate: 0.1 - 40℃/min.
Precision of the Temperature Control: ±0.1℃
Temperature Precision: 1%
5. Sample Injection
Capilliary Column, Packed Column, Cold-headed Column

Split/Splitless Injectors

 

 