Chameleon Multi-label Mikroplaka Okuyucu

Chameleon Multi-label Mikroplaka Okuyucu

Up to five technologies in one unit!
Read absorbance, luminescence, liquid scintillation, fluorescence intensity (FI), and time-resolved fluorescence (TRF)
Designed to read 24-, 96- and 384-well plates

mikroplaka okuyucu6 

These microplate readers are compact in size and save valuable bench space. An internal microprocessor with easy-to-use software accesses all instrument functions. Choose between single or combined function units. All models can have additional functions added at a later timeFor advanced data analysis order optional DeltaSoft PC software. DeltaSoft software includes nine curve fitting programs, statistical analysis of sample concentration calculations, detailed kinetic analysis and report templates that are ExcelTM compatible.

The Liquid Scintillation Counter has preset modes for the most commonly used radioisotopes (3H, 14C, 35S, 33P and 125I). Use for scintillation proximity assays (SPA), flash plate assays, cell proliferation studies, radioimmunoassays (RIA) and immunoradiometric assays (IRMA).

The Fluorescence Reader is ideal for both short life-time fluorophores and long life-time fluorescence using Fluorescence Intensity (FI), and Time-Resolved Fluorescence (TRF) modes. Applications include DNA hybridization studies, binding studies, enzyme assays, ligand receptor binding assays, cytotoxicity and immunoassays.
The Luminometer measures both �glow� and �flash� type reagents. Use for ATP detection, reporter gene assays, luminescence immunoassays, and toxicology. Luminescence function requires an injector (sold separately).
The Absorbance Reader has a wavelength range from 240 to 1000 nm. Typical applications include DNA quantification, colorimetric assays and enzyme assays.

Specifications:

Detectors: photon counting photomultiplier and photo diode
Optical type: direct illumination (fiberless optics)
Light source: xenon flash lamp-closed loop
Fluorescence sensitivity:
Fluorescence intensity (FI): <2 fmol
Fluorescence polarization (FP): 5mP
Time-resolved fluorometry (TRF): <100amol Eu
Excitation: 230-620 nm
Emission: 360-630 nm (700nm optional)
Dynamic range: >5 decades
Luminescence sensitivity: <1 amol ATP/well, >6 decades dynamic range
Liquid Scintillation Counting Efficiency: 30% for 3-H, 50% for I-125, and 80% for C-14 in a 96-well plate
Absorbance::
Linearity: 2%
Dynamic range: 3.0 OD
Wavelength range: 220 to 900nm


Specifications
Function Liquid scintillation and luminescence
Detector photon counting photomultiplier and photo diode
Optics direct illumination (fiberless optics)
Light source xenon flash lamp-closed loop
Luminescence sensitivity 1 amol ATP/well
dynamic range >6 decades dynamic range
Liquid scintillation counting 30% for 3H in a 96-well plate
Efficiency 50% for 125I and 80% for14C in a 96-well plate
Output RS-232
Power 110 to 240 VAC, 50/60 Hz
Dimensions 15-1/2"L x 14-1/2"W x 11-3/4"H