SUNDY SDS516 Sulfur Analyzer

SDS516 Sulfur Analyzer

  

HIGHLIGHTS

Coulomb Sulfur Analyzer working principles:
A sample combusts and decomposes in air current with catalyst, the sulfur in coal produces sulfide in the reaction, and among the products, the sulfur dioxide absorbed by potassium iodide (KI) solution . Titrate the iodine produced by electrolyzing potassium iodide, and calculate the total sulfur content in coal according to quantity of electricity consumed when electrolyzing.
The total sulfur content, St, ad, of the sample, expressed as a percentage by mass, is given by the equation below:
St, ad=m1/m×100
Where
St, ad is total sulfur content in general analysis samples, expressed as a percentage by mass, %;
m1 is displayed in coulomb integrator, expressed in milligram, mg;
m2 is the mass of the sample, expressed in milligram, mg.

  • With chain type automatic sample sending device which allows adding and replacing samples during tests, the instrument could accept 17 samples at once. All samples’ introducing, testing and discarding are automatic and completed continuously.
  • Horizontal furnace with a super wide uniform temperature zone(≥90mm) ensures complete combustion of coal samples.
  • Equipped with dilated and thickened electrolysis electrodes and a unique positioning slot which ensures that the electrolytic cell is fixed at the same position every time, the SDS616 can stir the electrolyte more evenly and then electrolyze the gases more completely.
  • Gas leak check is convenient and easy.
  • With CAN bus, several calorimeters can be controlled by a single PC without mutual influence.
  • Balance could be connected and data could be transmitted by real-time network transmission.

  

SPECIFICATIONS

Sulfur Determination Range: (0.01~35) %
Temperature Control Precision: ±3℃
Sulfur Resolution: 0.01%
Furnace Temperature: 1150℃ for coal, 900℃ for oil
Sample size: (8-110) mg
Analysis time: 3~6min /sample
Autoloader: 17 samples
Power Requirement: 220 V(-15%-10%),50Hz
Max Power: 3kw

  

APPLICATIONS AND CONFORMANCE WITH STANDARDS

Applications

It can be used to measure the total sulfur content in combustibles, such as coal, coke, petroleum and other solid biomass fuels. 

Conformance with Standards:

ISO351 Solid mineral fuels-Determination of total sulfur-High temperature combustion method
ASTM D 4239-2008 Standard Test Methods for Sulfur in the Analysis Sample of Coal and Coke Using High-Temperature Tube Furnace Combustion Methods1
GB/T214-2007 Total Sulfur Analysis in Coal  

« C / H / N - Sulfur Analyzers