The BINDER environmental simulation chamber of the MK series is suitable for heat or cold testing between -40 °C and 180 °C. The APT.line™ preheating chamber technology uniquely simulates a natural environment. For cyclical temperature testing, this environmental simulation chamber is a smart alternative to complex individual solutions.

Advantages:
  • State-of-the-art reliability
  • User-friendly chamber interior
  • Comprehensive standard equipment

 

Areas of application:
  • Automotive
  • Electronics / Semiconductor Industry
  • Plastics Industry

 

FeaturesCustomer benefitsCharacteristics
APT.line™ climate technology
  • Same test conditions throughout the chamber interior
  • Independent of specimen size and quantity

APT.line™

  • Uniform circulation even under full load
  • Homogeneous climate conditions throughout test specimens
Standard equipment
  • Very good price/performance ratio

Well equipped

  • Heated viewing window
  • LED illumination
  • Rugged chassis with rollers from 115 liters
  • Ethernet interface
Unit design
  • Minimum space requirements
  • Convenient, safe access
  • Easy assembly

Good use of space

  • Optimal ratio of usable space and footprint
  • All operator controls accessible from the front
  • Wide construction
Production
  • Reliable devices with long service lives
  • Short delivery times
  • Premium quality made in Germany
  • Highly automated series production (20,000 units per year)
  • High-quality materials, state-of-the-art production technology
Accessories and Services
  • Complete system from one source

Comprehensive product portfolio

  • Additional production lines with drying and vacuum chambers
  • Control and documentation software APT-COM™
  • BINDER Data Logger Kits
  • Water treatment with BINDER PURE AQUA SERVICE
  • Years of proven and recognized validation and documentation materials

 

Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results

  • Temperature range of -40 °C to 180 °C
  • MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
    • User friendly LCD screen
    • Easy-to-read menu guide
    • Integrated electronic chart recorder
    • Variety of options for the graphic display of process parameters
    • Real time clock
  • Adjustable ramp function via program editor
  • Access port Ø 80 mm, top
  • Heated viewing window with LED interior lighting
  • Temperature safety device class 2 (DIN 12880) with visual and audible temperature alarm
  • RS 422 interface for communication software APT-COM™ DataControlSystem
  • 1 stainless steel rack
  • BINDER test confirmation
  • BINDER Communication software APT-COM™ 3 Basic Edition
  •  

    MK 53 (E2.1)
      Exterior dimensions
    Width (incl. 80 mm access port with plug) (mm) 745
    Height (incl. feet) (mm) 1245
    Depth (incl. door handle, I-triangle, connection 55 mm) (mm) 795
    Wall clearance, side (mm) 160
    Wall clearance, rear (mm) 100
    Viewing window width (mm) 280
    Viewing window height (mm) 280
    Number of doors (ea.) 1
    Interior dimensions
    Width (mm) 402
    Height (mm) 402
    Depth (mm) 330
    Interior volume (l) 53
    Shelves (number standard/max.) 1 / 5
    Load per shelf (kg) 15
    Permitted total load (kg) 40
    Weight (empty) (kg) 150
    Temperature data
    Temperature range (°C) -40 - 180
    Temperature variation (± K) 0,4 - 2,0
    Temperature fluctuation (± K) 0,1 - 0,5
    Recovery time after door was open for 30 sec.
    at -10 °C (min.) 5
    at 70 °C (min.) 1
    at 150 °C (min.) 5
    Mean warm-up rate acc. to factory standard (°C/min.) -40 °C to 180 °C 4,6
    Mean cooling rate acc. factory standard (°C/min.) 180 °C to -40 °C 4,1
    Heat compensation, max. (W) 500
    Electrical data
    IP protection class acc. to EN 50529 IP 20
    Voltage (± 10 %) 50 Hz (V) 230, 1 N ~
    Nominal power (kW) 2,6
    Energy consumption at 20 °C (kW) 1) 1,02
    Noise level (approx. dB(A)) 59

     

    1) These values can be used for dimensioning air condition systems.

    MK 115 (E3.1)
      Exterior dimensions
    Width (incl. 18 mm access port with plug) (mm) 1000
    Height (incl. casters) (mm) 1725
    Depth (incl. door handle, I-triangle, connection 55 mm) (mm) 915
    Wall clearance, side (mm) 200
    Wall clearance, rear (mm) 300
    Viewing window width (mm) 288
    Viewing window height (mm) 222
    Number of doors (ea.) 1
    Interior dimensions
    Width (mm) 600
    Height (mm) 480
    Depth (mm) 400
    Interior volume (l) 115
    Racks (number standard/max.) 1 / 4
    Load per shelf (kg) 30
    Permitted total load (kg) 60
    Weight (empty) (kg) 260
    Temperature data
    Temperature range (°C) -40 - 180
    Temperature variation (± K) 0,1 - 2,0
    Temperature fluctuation (± K) 0,1 - 0,5
    Mean warm-up rate acc. to IEC 60068-3-5 (°C/min.) 5,3
    Mean cooling rate acc. to IEC 60068-3-5 (°C/min.) 5,0
    Heat compensation, max. (W) 2000
    Electrical data
    IP protection class acc. to EN 50529 IP 20
    Voltage (± 10 %) 50 Hz (V) 400, 3 N ~
    Nominal power (kW) 3,5
    Energy consumption at 20 °C (Wh/h) 1) 650
    Noise level (approx. dB(A)) 62

     

    1) These values can be used for dimensioning air condition systems.

    MK 240 (E3.1)
      Exterior dimensions
    Width (incl. 18 mm access port with plug) (mm) 1135
    Height (incl. casters) (mm) 1715
    Depth (incl. door handle, I-triangle, connection 55 mm) (mm) 1000
    Wall clearance, side (mm) 300
    Wall clearance, rear (mm) 200
    Viewing window width (mm) 508
    Viewing window height (mm) 300
    Number of doors (ea.) 1
    Interior dimensions
    Width (mm) 735
    Height (mm) 700
    Depth (mm) 443
    Interior volume (l) 228
    Racks (number standard/max.) 1/ 6
    Load per shelf (kg) 30
    Permitted total load (kg) 70
    Weight (empty) (kg) 360
    Temperature data
    Temperature range (°C) -40 - 180
    Temperature variation (± K) 0,1 - 1,2
    Temperature fluctuation (± K) 0,1 - 0,5
    Mean warm-up rate acc. to IEC 60068-3-5 (°C/min.) 5
    Mean cooling rate acc. to IEC 60068-3-5 (°C/min.) 4.5
    Heat compensation, max. (W) 2000
    Electrical data
    IP protection class acc. to EN 50529 IP 20
    Voltage 50 / 60 Hz (V) 400, 3 N ~
    Nominal power (kW) 4,2
    Energy consumption at 20 °C (W) 1) 1300
    Noise level (approx. dB(A)) 62

     

    1) These values can be used for dimensioning air condition systems.

    MK 720 (E3.1)
      Exterior dimensions
    Width (incl. 80 mm access port with plug) (mm) 1615
    Height (incl. casters) (mm) 2005
    Depth (incl. door handle, I-triangle, connection 55 mm) (mm) 1230
    Wall clearance, rear (mm) 300
    Wall clearance, side (mm) 200
    Viewing window width (mm) 508
    Viewing window height (mm) 300
    Number of doors 1
    Interior dimensions
    Width (mm) 1200
    Height (mm) 1020
    Depth (mm) 600
    Interior volume (l) 734
    Shelves (number standard/max.) 1 / 11
    Load per shelf (kg) 40
    Permitted total load (kg) 160
    Weight (empty) (kg) 570
    Temperature data
    Temperature range (°C) -40 - 180
    Temperature variation (± K) 0,3 - 2,0
    Temperature fluctuation (± K) 0,1 - 0,5
    Mean warm-up rate acc. to IEC 60068-3-5 (°C/min.) 4
    Mean cooling rate acc. to IEC 60068-3-5 (°C/min.) 4,5
    Heat compensation (W) 6500
    Electrical data
    Voltage (± 10 %) 50 Hz (V) 400, 3 N ~
    Nominal power (kW) 7,2
    Energy consumption at 20 °C (W) 1) 1900
    Noise level (approx. dB(A)) 65

     

    1) These values can be used for dimensioning air condition systems.

    OPTIONS
    Access ports with silicone plug 30, 50, 80, 100, 125 mm O
    Securing elements for additional fastening of racks (1 set of 4 pieces) O
    Keyboard lock O
    Analog output for temperature 4 - 20 mA with 6-pin DIN socket (output not adjustable) O
    Factory calibration certificate, measurement in center of chamber at 150 °C or at specified testing temperature O
    Extension to factory calibration certificate. Each additional measurement at additional measuring point or testing temperature O
    Additional measuring channel for digital display of specimen temperature with flexible PT 100 temperature sensor, measured data recorded via unit interface O
    Data Logger Kit T 220: For continuous temperature recording of -90 °C to 220 °C. Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for the connection at the BINDER unit O
    Data Logger Software: Configuration and evaluation software for all BINDER Data Logger Kits, incl. data cable (RS 232) O
    Rack, stainless steel O
    Reinforced rack, stainless steel, with 1 set of fasteners (4 pieces), max. load 70 kg O
    Shelf perforated, stainless steel O
    Notch-type access port in door, 100 x 35 mm O
    Temperature safety device for over and under temperature, Class 2 O
    Zero-voltage relay outputs accessible via 6-pin DIN socket. Additional module for controlling 3 relay outputs via 3 of the programmable controller's contacts O
    RS 422 interface O
    Locking door handle with key O