BINDER MDL Series Safety Drying Oven ( MDL115)

BINDER MDL Series Safety Drying Oven ( MDL115)

 

  

Safety drying oven with expended temperature range

The BINDER MDL 115 with enhanced control function works at temperatures up to 350 °C. This safety drying oven meets all EN 1539 requirements 1539 and is perfectly suited for high-temperature testing such as in the coil coating area.

Advantages:
  • Defined solvent quantity according to EN 1539
  • Wide temperature range up to 350 °C

 

Areas of application:
  • Chemicals
  • Surface Technology

 

 

FeaturesCustomer benefitsCharacteristics
APT.line™
  • Same test conditions throughout the chamber interior
  • Independent of specimen size and quantity

APT.line™ preheating chamber for maximum precision

  • Uniform circulation throughout the inner chamber
  • Homogeneous temperature conditions throughout test specimens
Safety concept
  • Defined solvent quantity
  • Safe standard-compliant drying
  • Complies with EN 1539:
    • Fresh air monitoring with automatic switchoff
    • Audible and visual alarm
    • Class 2 temperature limiter
  • Safety class IP 33
  • Easy to replace fresh air filter
Performance feature
  • Short warm up times
  • Broad range of applications, e.g. coil coating tests
  • High air exchange rate
  • Monitored air exchange volume
  • Large power reserves up to 9 kW
  • Temperature range up to 350 °C
Inner chamber concept
  • Maximum occupational safety
  • Easy loading and unloading of specimen material
  • Easy to clean
  • Very tight door closure with 2-point door latch
  • Low heat dissipation due to 60 mm insulation
  • Rack with tilt protection
  • Complete inner chamber made of stainless steel
  • No permanent fixtures
  • Silicone free
  • Viton (FKM) door seal
Controller
  • Convenient HMI (Human Machine Interface)
  • Convenient documentation
  • All measured values read at a glance
  • Controller with color display and 25 storable programs
  • Integrated digital continuous-line recorder for monitoring limits and alarm function
Quality
  • Reliable devices with long service lives
  • Short delivery times
  • Minimal maintenance and operating costs
  • Premium quality made in Germany
  • Highly automated series production (20,000 units per year)
  • High-quality materials, state-of-the-art production technology
Unit options and service
  • Optimal solution for numerous applications
  • BINDER INDIVIDUAL for customer-specific solutions
  • Worldwide BINDER Service

Comprehensive product portfolio:

  • Various options: coil coating extension, temperature measurement of the specimen, calibration certificate
  • Control and documentation software APT-COM™
  • Drying ovens, vacuum drying ovens, climate chambers
  • Worldwide service network

 

Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results

  • Temperature range from 5 °C above ambient temperature to 350 °C
  • All safety features met according to EN 1539
  • Heating load 9.0 kW
  • Door gasket made of of high temperature resistant silicone
  • Rear exhaust connector Ø 100 mm
  • MCS controller with 25 storable programs of 100 sections each for a maximum of 500 program segments
  • User-friendly LCD screen
    • Easy-to-read menu guide
    • Integrated electronic chart recorder
    • Variety of options for the graphic display of process parameters
    • Real-time clock
  • Replacable fresh-air filter cartridge, class F6 (EU6-fine particle filter for particle sizes between 1 ... 10 µm)
  • Independent adjustable temperature safety device class 2 (DIN 12880), with acoustic and visual temperature alarm
  • Fresh-air monitoring with audible and visual alarm and automatic deactivation of heating
  • RS 422 interface for communication software APT-COM™ DataControlSystem
  • 2 chrome-plated racks included
  • BINDER test confirmation
  •  

    MDL 115
    Exterior dimensions
    Width (mm) 835
    Height (incl. feet) (mm) 800
    Depth (plus door handle 50 mm) (mm) 685
    Wall clearance, rear (mm) 100
    Wall clearance, side (mm) 100
    Exhaust duct (outer Ø mm) 100
    Steam space volume (l) 156
    Interior dimensions
    Width (mm) 600
    Height (mm) 435
    Depth (mm) 435
    Interior volume (l) 115
    Racks (number standard/max.) 2 / 5
    Load per rack (kg) 20
    Permitted total load (kg) 50
    Weight (empty) (kg) 90
    Temperature range
    Temperature range approx. 5 °C above ambient temperature to (°C) 350
    Temperature variation
    at 70 °C (± K) 2
    at 150 °C (± K) 3,4
    at 300 °C (± K) 7
    Temperature variation, extension in the door
    at 70 °C (± K) 2
    at 150 °C (± K) 3
    at 300 °C (± K) 8
    Temperature fluctuation (± K) 0,5
    Warm-up time from 2)
    to 70 °C (min.) 3,5
    to 150 °C (min.) 6
    to 300 °C (min.) 10
    Recovery time after doors were open for 30 sec. 2)
    at 70 °C (min.) 0,5
    at 150 °C (min.) 2
    at 300 °C (min.) 4
    Recovery time after door was open for 30 sec. 2), extension in the door
    at 70 °C (min.) 0,5
    at 150 °C (min.) 1
    at 300 °C (min.) 2
    Ventilation (approx. x/min.) 3
    Air circulation (approx. x/min.) 40
    Exhaust air volume flow (approx. l/min. m³/h) 400 (24,0)
    Air flow velocity (m/sec) 0,8 - 1,2
    Maximum amount of solvent (g) (at T-180 °C, M-100 g/mol, U-40 g/m3, K = 0,5) 6,65
    Electrical data
    IP protection class acc. to EN 60529 IP 33
    Voltage (± 10%) 50 / 60 Hz (V) 400 V / 3N~
    Nominal power (kW) 9,0
    Energy consumption at 150 °C (W) 1130

    1)
    T = drying temperature
    M = molecular mass
    U = lower explosion limit
    K = solvent vapor concentration as percentage at lower explosion limit

    2) To 98% of the set value

     

    OPTIONS
    Rack,chrome-plated or stainless steel O
    Shelf, perforated, stainless steel O
    Locking door handle with key O
    Keyboard lock O
    Base on castors O
    Door flap for coil-coating tests O
    Replacement air filter (class F6/EU6 – fine particle filter for particle sizes between 1 µm and 10 µm), 100 × 520 × 22 mm, with aluminum frame O
    Additional measuring channel for digital display of specimen temperature, with clip temperature sensor. Measured data recorded via RS 422 interface O
    Temperature measurement acc. to DIN 12880 (27 measuring points) at 150 °C or at specified temperature with measuring protocol and certificate O
    Factory calibration certificate. Measurement in center of chamber at 150 °C (302 °F) or at specified testing temperature O
    Extension to factory calibration certificate. Each additional measurement at an additional measuring point or temperature O