- Industrial
- Automatic Paper Testing ( Autoline Profiler )
- Color Measurement Instruments
- Electro Analyzer
- Gas Analyzers
- Industrial Textile Treating Systems
- On-Line Chemical Analyzers
- Package Quality Testing Instruments
- Paper & Board Process Optimization Instruments
- Paper Testing Instruments
- Petroleum Testing Instruments
- Pulp Testing Instruments
- Sieve Shakers / Sieves
- Laboratory
- Amino Acid Analyzers
- Autoclaves
- Balances
- Baths and Circulators
- Bio Reactors / Fermentors
- Blenders
- Block Heaters
- Calibration Equipment
- Calorimeters
- Centrifuges
- Chromatography
- Conductivity Meters
- Deep Freezers / Freezers
- Electrical Test Equipment
- Electronic Test Instruments
- Electrophoresis and Accessories
- Flowmeters
- Fraction Collectors
- Furnaces
- Glassware Washers
- Heating Mantles
- Homogenizers
- Hot Plates
- Incubators
- Ion Meters
- Liyofilization (Freeze Drying) Instruments
- Magnetic Stirrers
- Melting Point Apparatus
- Microplate Readers
- Microscopes
- Mills
- Mixers
- Moisture Analyzers
- Organic Synthesis Systems
- Ovens
- Oxygen Meters
- pH Meters
- Photometers
- Polorimeters
- Presses
- Pressure / Vacuum Meters
- Pumps
- Reactors
- Recorder and Dataloggers
- Refractometers
- Rotary Evaporators
- Salinity Meters
- Sample Preparation Bombs
- Shakers
- Spectrophotometers
- Sterilizers
- Temperature Controlled Heating & Cooling Systems
- Temperature Measuring Instruments
- Thermocycler (PCR)
- Thermohygrometers
- Ultrasonic Cleaners
- Vibration / Leak Detection Instruments
- Viscometers
- Water Purification Systems
- Water Testing Instruments
AB Lorentzen & Wettre Paper Brightness /Opacity / Colour Testing
AB Lorentzen & Wettre Paper Brightness /Opacity / Colour Testing
Elrepho – fast and accurate colour measurement
Elrepho is the paper industry's own spectrophotometer. It measures colour, brightness, opacity and whiteness of paper, paperboard, pulp, coating inks and fillers. The measurement conforms to all established standards for optical measurements, i.e. SCAN, DIN, ISO, TAPPI, CPPA and AFNOR. Important details such as cut-off filters and measurement aperture also comply with ISO 2469. The instrument is simple to operate with measurement table, and easy-to-use software.
L&W Colour-Brightness software, is optimized for shift testing and has complete user interfaces for identification, measurement, and reporting. Grades with limit values and target values can be set as required. The software reports most optical characteristics. Calculations can also be implemented to meet the user’s needs.
Traceable calibration
Elrepho comes with traceable calibration in accordance with the ISO’s hierarchy for photometric calibration and UV calibration at level C and D65. Working standards and a delivery certificate are included.
The software has an easy function for checking the calibration, or making a new calibration, when necessary. A large number of reference and working standards, with target values can be stored.
Fast and easy handling of samples
Elrepho is designed to make laboratory work easier. A sample is placed on the measurement table, and then automatically positioned against the instrument’s measurement aperture. For opacity measurements, the black cavity is controlled automatically. In fact, the measurement software controls everything.
Simple measurement sequences
The software automatically measures according to the conditions specified by the user, e.g. D65 illuminant, C illuminant, or a 420 nm cut-off filter. All at the touch of a button.
Measurement results can be copied to other programs by a simple copy and paste function. Alternatively, data can be saved in an Excel-compatible file. Sample identities can be retrieved from (and measurement results sent to) other databases using standard commands and tools.
Graphic presentations
The program features a number of ways to display the results graphically. The trend diagram is used to monitor the production over a longer period, while the L*a*b* plot shows the colour coordinates for a series of samples. The reflectance curve provides important information about colour dyeing and the fluorescence of the sample.
Measures effective residual ink concentration, ”ERIC”, of recycled pulp and paper
The measurement is made on a sheet and opaque pad according to the ISO 22 754 and TAPPI T 567 standards respectively. This measurement can alternatively be made on an opaque pad only. This latter method is not standardized. An alternative is to measure the ink elimination between two process stages with the INGEDE method.
A complete system
The delivery includes Elrepho spectrophotometer, L&W Colour-Brightness software, PC, printer and the accessories necessary for checking the calibration. Elrepho comes in a standard version with a measurement table for fast sample handling, but can also be supplied without the measurement table.
Specifications- Optimized measurement program for routine testing
- Reliable measurement even on small surfaces
- One calibration for all filter conditions
- Fast measurement sequences with different UV settings
- Residual ink evaluation
- Measures Whiteness and Brightness at D65 and C illuminant
|
Specification
|
|
|
Elrepho – Code 071 (with measurement table), Code 071E (with measurement table, with ERIC)
Code 070 (without measurement table, without ERIC), Code 070E (without measurement table, with ERIC) |
|
|
Inclusive
|
L&W Colour-Brightness software, PC, laser printer, working standards, apertures, black cavity, cables
|
|
Alternative
|
Same as above, without PC/printer
|
|
Measurement
|
|
|
Photometric range
|
0–200 %
|
|
Repeatability
|
< 0.01 ΔE CIELAB units on white ceramic tiles
|
|
Reproducibility
|
< 0.25 ΔE CIELAB units (average) on NCS colour samples
|
|
Instrument
|
|
|
Spectral range
|
360–700 nm
|
|
Reporting range
|
400–700 nm adjustable to 360–700 nm
|
|
Measurement geometry
|
d/0º, according to ISO 2469
|
|
Measuring principle
|
SP 2000 dual-beam monochromator with 2 × 256 diodes in 2 nm intervals
|
|
Apertures
|
Standard 34 mm. Alternatively 9 mm or 6.6 mm. Automatic focus on the test area with 30 mm, 5 mm or 2.5 mm diameter. Optional 18 mm aperture with 16 mm test area diameter
|
|
Bandwidth
|
10 nm
|
|
Light source
|
Pulsed xenon lamp, D 65 – filtered
|
|
Measurement time
|
4–20 s, depending on the type of measurement
|
|
Filters
|
395 nm and 420 nm (UV cut-off), automatically selected from the program
|
|
UV level
|
D65 and C, automatically controlled from the program after calibration
|
|
Software
|
|
|
Operating system
|
Microsoft Windows
|
|
User program
|
L&W Colour-Brightness measurement program specially developed for Elrepho
|
|
Functions
|
Overview, control, identification, measurement, reporting, data output
|
|
Help functions
|
Calibration, calibration control, grade settings, calculated properties
|
|
Illuminant/Observer
|
D65/10°, C/2°, A/10°, D50/2°, T11/10°
|
|
Calibration
|
Commands for photometric and UV calibration. Calibrations and working standards with target values can be stored
|
|
Data output
|
L&W Autoline and Excel-compatible
|
|
Tabular reports
|
Test report
|
|
Graphical reports
|
Reflectance factor, opacity, scattering, absorption, trend, L*-a*-b*
|
|
Properties
|
X, Y, Z, L*, a*, b*, x, y, Rx, Ry, Rz, CIE Whiteness/Tint-D65 resp. C, Fluorescence, ISO Brightness, D65-Brightness, Yellowness, Opacity, Transparency, K/N-value, IE, Residual ink, DW, pe, s, k, sR457, kR457, Metamerism, ΔE and others
|
|
PC
|
1.2 GHz or higher, Windows operating system, 2000, XP or later version, 17-in screen, serial port
|
|
Printer
|
Laser or ink-jet printer
|
|
Installation requirements
|
|
|
Power
|
400 W
|
|
Instrument air
|
> 0,5 MPa (70 psi) (not 070)
|
|
Air consumption
|
0,2 m³/h (0.1 ft³/min) NTP (not 070)
|
Size
|
Dimensions
|
|
|
Volume
|
|
Instrument
|
0.4 × 0.5 × 0.6 m
16 × 20 × 24 in |
|
0.4 m³
14 ft³ |
|
Printer
|
0.3 × 0.3 × 0.3 m
12 × 12 × 12 in |
|
0.2 m³
7 ft³ |
|
PC/screen
|
0.6 × 0.6 × 0.5 m
24 × 24 × 20 in |
|
0.4 m³
14 ft³ |
|
|
|
|
|
|
Net weight
|
26 kg
57 lb |
Gross weight
|
50 kg
110 lb |
|
PC/printer/screen
|
41 kg
90 lb |
|
61 kg
134 lb |
Standards
|
Applicable standards
|
|
PAPTAC E1, E2, DIN 53 145/53 146/53 147/54 500, ISO 2469/2470/2471/5631/9416/11475/11476/22754/22891/, SCAN G5/P3/P8/P66/P71/P 72, TAPPI T 519/525/527/560/567
|
« Paper Testing Instruments