METTLER TOLEDO Digital Melting Point Apparatus

METTLER TOLEDO Digital Melting Point Apparatus

Precision accuracy of 0.5C provides high reproducibility of results
Meet GLP requirements with optional printer

Digital display eliminates interpretation errors

er1

This automatic melting point apparatus is ideal for determining sample purities for quality control, educational, and research and development applications. Unit measures the reflection of incident light in the crystalline state and the increase in light transmittance during the melting process. Operation is simpleinsert the test tube containing one to three mg of sample into the instrument, select the starting temperature and heating rate (user selectable between 0.1C and 10C per minute in steps of 0.1C), and the visual and audible signals will indicate when the melting point is acheived.
The operator interface provides graphic display aids making set up and programming simple. The large LCD displays temperature readings to 0.1C. Internal fan allows for rapid cooling to user defined temperature settings. RS-232 interface provides connection to printer (order below) for further data analysis and for GLP applications.

Specifications
Temp range ambient to 300°C
Resolution 0.1°C
Temp accuracy Ambient to 200°C: ±0.5°C|200 to 300°C: ±0.5 to 0.8°C (linear increase)
Heat-up time 50 to 300°C, 3 minutes (0.1 to 10°C/minute)
Display 4-digit LCD, 1/2"H
Power VAC 100 to 240
Hz 50/60
watts 73
Dimensions 9-1/2"W x 4-3/4"H x 11-1/4"D
Reproducibility ±0.1°C

 